Omron SYSMAC FH Series User Manual page 268

For communication settings
Hide thumbs Also See for SYSMAC FH Series:
Table of Contents

Advertisement

2 Methods for Connecting and Communicating with External Devices
Setting item
Timeout [s]
Output data
size
User area
Precautions for Correct Use
About Output data size and User Area
All line's Output data size and User Area are necessary to be equal when you select the Multi-
line random trigger in Operation mode.
4
Click Apply to apply the settings.
Click Close to close the System Settings dialog box.
2 - 206
Setting value
[Factory default]
0.5 to 120.0
[10.0]
Result Data Format 0
(32 bytes)
Result Data Format 1
(64 bytes)
Result Data Format 2
(128 bytes)
Result Data Format 3
(256 bytes)
[None]
ON
FH/FHV Series Vision System User's manual for Communication Settings (Z342-E1)
Description
Valid only when Output control is set to
Handshaking.
A timeout error occurs when no response from ex-
ternal devices is received at the following timing
within the time that has been set.
In the following cases, a timeout error occurs
when the state of each signal does not change
within the time that has been set.
If the DSA signal were not turned ON even after
a certain time passed from when measurement
has been completed.
The DSA signal turns OFF after the GATE flag
has turned ON.
The DSA signal turns ON after the GATE flag
has turned OFF.
Set the data size to output as measurement re-
sults for each line. The settings are reflected at the
restart after they were stored.
There are four types in the output data size: 32,
64, 128, and 256 bytes.
Result Data Format 0 (32 bytes)
Out put data 0 to 7 of 4 bytes can be used and to-
tal size of the output data is 32 bytes.
Result Data Format 1 (64 bytes)
Out put data 0 to 15 of 4 bytes can be used and
total size of the output data is 64 bytes.
Result Data Format 2 (128 bytes)
Out put data 0 to 31 of 4 bytes can be used and
total size of the output data is 128 bytes.
Result Data Format 3 (256 bytes)
Out put data 0 to 63 of 4 bytes can be used and
total size of the output data is 256 bytes.
Set whether or not to use the User Area (user in-
put and output areas)
Data type of User Input Area 0 to 3 is DINT. Da-
ta type of User Input Area 4 to 5 is LREAL.
Data type of User Output Area 0 to 3 is DINT.
Data type of User Output Area 4 to 5 is LREAL.