Figure 2-29 Connections for the Second Portion of Conversion Compression
Measurement
9. To set the frequency offset mode LO frequency, press:
System
INSTRUMENT MODE
LO MENU
FREQUENCY:CW
10.To select the converter type, press:
RETURN
UP CONVERTER
11.To select a low-side LO measurement configuration, press:
RF>LO
FREQ OFFS ON
In this low-side LO, up converter measurement, the analyzer source frequency is offset
lower than the receiver frequency. The analyzer source frequency can be determined from
the following equation:
receiver frequency (800 MHz) − LO frequency (600 MHz) = 200 MHz
The measurements setup diagram is shown in
Conversion Compression Using the Frequency Offset Mode
FREQ OFFS MENU
600
M/µ
Figure
Making Mixer Measurements
2-30.
2-41