Operating Concepts

S-Parameters

S-Parameters
The
key accesses the S-parameter menu which contains softkeys that can be used
Meas
to select the parameters or inputs that define the type of measurement being performed.

Understanding S-Parameters

S-parameters (scattering parameters) are a convention used to characterize the way a
device modifies signal flow. A brief explanation of the S-parameters of a two-port device is
provided, however, for additional details, refer to Application Notes 95-1 and 154.
S-parameters are always a ratio of two complex (magnitude and phase) quantities.
S-parameter notation identifies these quantities using the numbering convention:
S out in
where the first number (out) refers to the test-device port where the signal is emerging and
the second number (in) is the test-device port where the signal is incident. For example,
the S-parameter S
identifies the measurement as the complex ratio of the signal
21
emerging at the test device's port 2 to the signal incident at the test device's port 1.
Figure 7-3
is a representation of the S-parameters of a two-port device, together with an
equivalent flowgraph. In the illustration, "a" represents the signal entering the device and
"b" represents the signal emerging. Note that a and b are not related to the A and B input
ports on the analyzer.
7-21