Agilent Technologies 8753ET User Manual page 145

Network analyzers
Hide thumbs Also See for 8753ET:
Table of Contents

Advertisement

Making Mixer Measurements
High Dynamic Range Swept RF/IF Conversion Loss
Figure 2-20
shows the conversion loss of this low-side LO mixer with output filtering.
Notice that the dynamic range from the pass band to the noise floor is well above the
dynamic range limit of the R Channel. If the mixer under test also contained
amplification, then this dynamic range would have been even greater due to the
conversion gain of the mixer.
Figure 2-20 Example of Swept IF Conversion Loss Measurement