Calibrate And Measure - Agilent Technologies 8753ET User Manual

Network analyzers
Hide thumbs Also See for 8753ET:
Table of Contents

Advertisement

MORE
SEGMENT POWER
SEGMENT IF BW
RETURN
DONE
9. Press
DONE LIST FREQ [SWEPT]

Calibrate and Measure

1. Remove the DUT and perform a full two-port calibration. Refer to
"Calibrating for Increased Measurement Accuracy."
2. With the thru connected, set the scale to autoscale to observe the benefits of using
swept list mode.
• The segments used to measure the stopbands have less noise, thus maximizing
dynamic range within the stopband frequencies.
• The segment used to measure the passband has been set up for faster sweep speed
with more measurement points.
Figure 1-58 Calibrated Swept List Thru Measurement
3. Reconnect the filter and adjust the scale to compare results with the first filter
measurement that used a linear sweep.
In
Figure
1-59, notice that the noise level has decreased over 10 dB, confirming that the
noise reduction techniques in the stopbands were successful. Also, notice that the
stopband noise in the third segment is slightly lower than in the first segment. This is
due to the narrower IF bandwidth of the third segment (300 Hz).
10
x1
300
x1
.
Making Measurements
Using the Swept List Mode to Test a Device
Chapter 6 ,
1-73

Advertisement

Table of Contents
loading

This manual is also suitable for:

8753es

Table of Contents