Figure 6-7 Standard Connections for Enhanced Response Calibration
8. To measure the standard, when the displayed trace has settled, press:
OPEN
The analyzer displays WAIT - MEASURING CAL STANDARD during the standard
measurement. The analyzer underlines the
standard.
9. Disconnect the open, and connect a short circuit to the test port.
10.To measure the device, when the displayed trace has settled, press:
SHORT
The analyzer measures the short circuit and underlines the
11.Disconnect the short, and connect an impedance-matched load to the test port.
12.To measure the standard, when the displayed trace has settled, press:
, select the type of load you are using, and then press
LOADS
the analyzer has finished measuring the load.
Notice that the
LOADS
13.To compute the reflection correction coefficients, press:
STANDARDS DONE
14.To start the transmission portion of the correction, press
15.Make a "thru" connection between the points where you will connect your device under
test as shown in
Figure
softkey is now underlined..
6-7.
Calibrating for Increased Measurement Accuracy
Enhanced Frequency Response Error Correction
softkey after it measures the
OPEN
SHORT
DONE: LOADS
TRANSMISSION
softkey.
when
.
6-23