Operating Concepts
Measurement Calibration
Next, a short circuit termination whose response is known to a very high degree is used to
establish another condition as shown in
Figure
7-31.
Figure 7-31 Short Circuit Termination
The open circuit gives the third independent condition. In order to accurately model the
phase variation with frequency due to fringing capacitance from the open connector, a
specially designed shielded open circuit is used for this step. (The open circuit capacitance
is different with each connector type.) Now the values for E
, directivity, E
, source
DF
SF
match, and E
, reflection frequency response, are computed and stored. See
Figure
7-32.
RF
Figure 7-32 Open Circuit Termination
This completes the calibration procedure for one port devices.
7-46