Making Measurements
Using the Swept List Mode to Test a Device
Set Up the Lower Stopband Parameters
3. To set up the segment for the lower stopband, press
ADD
START
650
STOP
880
M/µ
NUMBER of POINTS
4. To maximize the dynamic range in the stopband (increasing the incident power and
narrowing the IF bandwidth), press
MORE
LIST POWER ON off
LIST IF BW ON off
RETURN
DONE
Set Up the Passband Parameters
5. To set up the segment for the passband, press
ADD
CENTER
900
SPAN
40
M/µ
STEP SIZE
.2
6. To specify a lower power level and a wider IF bandwidth for the passband, press
MORE
SEGMENT POWER
SEGMENT IF BW
RETURN
DONE
Set Up the Upper Stopband Parameters
7. To set up the segment for the upper stopband, press
ADD
START
920
STOP
1150
M/µ
NUMBER of POINTS
8. To maximize the dynamic range in the stopband (increasing the incident power and
narrowing the IF bandwidth), press:
1-72
M/µ
51
x1
until ON is selected
until ON is selected
M/µ
M/µ
−10
x1
3700
x1
M/µ
51
x1
SEGMENT POWER
SEGMENT IF BW
1000
10
x1
x1