Making Mixer Measurements
Fixed IF Mixer Measurements
TITLE
POW:LEV 13DBM
PERIPHERAL HPIB ADDR
21x1
TITLE TO PERIPHERAL
TITLE
FREQ:MODE CW;CW 600MHZ;:FREQ:CW:STEP 100MHZ
TITLE TO PERIPHERAL
DO SEQUENCE
SEQUENCE 2
Sequence 2 Setup
The following sequence makes a series of measurements until all 26 CW measurements
are made and the loop counter value is equal to zero. This sequence includes:
• taking data
• incrementing the source frequencies
• decrementing the loop counter
• labeling the screen
1. Press the following keys on the analyzer to create sequence 2:
NOTE
To enter the following sequence commands that require titling, an external
keyboard may be used for convenience.
DONE SEQ MODIFY
Seq
Seq
NEW SEQ/MODIFY SEQ
Taking Data
SPECIAL FUNCTIONS
Seq
Sweep Setup
Incrementing the Source Frequencies
Display
MORE
Input as title: FREQ:CW UP
DONE
Seq
SPECIAL FUNCTIONS
TITLE TO PERIPHERAL
PERIPHERAL HPIB ADDR
2-30
SEQUENCE 2 SEQ2
WAIT x
TRIGGER MENU
MANUAL TRG ON POINT
TITLE
ERASE TITLE
PERIPHERAL HPIB ADDR
21
x1
.1
x1
TITLE TO PERIPHERAL
19
x1