Calibrating for Increased Measurement Accuracy
One-Port Reflection Error Correction
13.To compute the error coefficients, press:
DONE: 1-PORT CAL
The analyzer displays the corrected data trace. The analyzer also shows the notation
Cor to the left of the screen, indicating that the correction is switched on for this
channel.
NOTE
The open, short, and load could be measured in any order, and need not follow
the order in this example.
You can save or store the error correction to use for later measurements.
NOTE
Refer to
for procedures.
14.This completes the one-port correction for reflection measurements. You can connect
and measure your device under test.
6-28