Making Measurements
Using Test Sequencing
The following list is the commands entered in
Notice that for longer sequences, only a portion of the list can appear on the screen at
one time.
Start of Sequence
RECALL PRST STATE
Trans: FWD S21 (B/R)
LOG MAG
CENTER
134 M/u
SPAN
50 M/u
SCALE/DIV
AUTO SCALE
3. To change a command (for example, the span value from 50 MHz to 75 MHz), move the
cursor (
) next to the command that you wish to modify, press:
or
• If you use the
commands are actually performed when the cursor points to them. This feature
allows the sequence to be tested one command at a time.
• If you wish to scroll through the sequence without executing each line as you do so,
you can press the
4. To delete the current command (for example, span value), press:
5. To insert a new value (for example, 75 MHz), press:
75
M/µ
6. Press
DONE SEQ MODIFY
Seq

Clearing a Sequence from Memory

1. To enter the menu where you can clear a sequence from memory, press:
MORE
CLEAR SEQUENCE
Seq
2. To clear a sequence, press the softkey of the particular sequence.
1-88
key to move the cursor through the list of commands, the
key and scroll through the command list backwards.
to exit the modify (edit) mode.
"Creating a Sequence" on page
1-84.