Calibrating for Increased Measurement Accuracy

Calibrating for Noninsertable Devices

Calibrating for Noninsertable Devices
A test device having the same sex connector on both the input and output cannot be
connected directly into a transmission test configuration. Therefore, the device is
considered to be noninsertable, and one of the following calibration methods must be
performed:
Adapter Removal (ES Analyzers Only) on page 6-41
Matched Adapters on page 6-45
Modify the Cal Kit Thru Definition on page 6-46
Figure 6-12 Noninsertable Device