Channel Stimulus Coupling

COUPLED CH on OFF
COUPLED CH ON
(The inactive channel takes on the stimulus values of the active channel.)
In the stimulus coupled mode, the following parameters are coupled:
• frequency
• number of points
• source power
• number of groups
• IF bandwidth
• sweep time
• trigger type
• gating parameters
• sweep type
• power meter calibration
Coupling of stimulus values for the two channels is independent of
in the display menu and
COUPLED CH OFF
on. In this mode, the analyzer alternates between the two sets of stimulus values and
displays the measurement data of both channels.
toggles the channel coupling of stimulus values. With
(the preset condition), both channels have the same stimulus values.
MARKERS: UNCOUPLED
activates an alternate sweep function when dual channel display is
Channel Stimulus Coupling
DUAL CHAN on OFF
in the marker mode menu.