Calibrating for Increased Measurement Accuracy
Frequency Response and Isolation Error Corrections
5. If your calibration kit is different than the kit specified under the
press:
CAL KIT
SELECT CAL KIT
If your type of calibration kit is not listed in the displayed menu, refer to
Calibration Kits" on page
6. To select a response and isolation correction and to start with the response portion of
the calibration, press:
CALIBRATE MENU
7. Connect the short or open calibration standard to the port you selected for the test port
(PORT 1 for S
or PORT 2 for S
11
NOTE
Include any adapters that you will have in the device measurement. That is,
connect the standard device to the particular connector where you will
connect your device under test.
Figure 6-6 Standard Connections for a Response and Isolation Error
Correction for Reflection Measurements
8. To measure the standard, press:
or
SHORT
OPEN
If the calibration kit you selected has a choice between male and female calibration
standards, remember to select the sex that applies to the test port and not the standard.
6-20
(select your type of kit)
7-57.
RESPONSE & ISOL'N
on ES analyzers).
22
CAL KIT [ ]
RETURN
RESPONSE
softkey,
"Modifying