Measuring Gain And Reverse Isolation Simultaneously (Es Analyzers Only) - Agilent Technologies 8753ET User Manual

Network analyzers
Hide thumbs Also See for 8753ET:
Table of Contents

Advertisement

Making Measurements
Measuring Amplifiers
Measuring Gain and Reverse Isolation Simultaneously
(ES Analyzers Only)
Since an amplifier will have high gain in the forward direction and high isolation in the
reverse direction, the gain (S
Therefore, the power you apply to the input of the amplifier for the forward measurement
(S
) should be considerably lower than the power you apply to the output for the reverse
21
measurement (S
). By applying low power in the forward direction, you'll prevent the
12
amplifier from being saturated. A higher power in the reverse direction keeps noise from
being a factor in the measurement and accounts for any losses caused by attenuators or
couplers on the amplifier's output needed to lower the output power into the analyzer. The
following steps demonstrate the features that best accomplish these measurements.
1. Press
Sweep Setup
Coupling the channels allows you to have the same frequency range and calibration
applied to channel 1 and channel 2.
2. Press
PORT POWER [UNCOUPLED]
Power
Uncoupling the port power allows you to apply different power levels at each port. In
1-51, the port 1 power is set to −25 dBm for the gain measurement (S
Figure
port 2 power is set to 0 dBm for the reverse isolation measurement (S
3. Press
Chan 1
Meas
1.
4. Press
Chan 2
Meas
2.
5. Perform an error-correction and connect the amplifier to the network analyzer. Refer to
the
Chapter 5 , "Optimizing Measurement Results,"
6. Press
Display
DUAL | QUAD SETUP
You can view both measurements simultaneously by using the dual channel display
mode. Refer to
Figure
the displayed measurements will not be continually updated and the annotation tsH
will appear on the left side of the display. Refer to
Protection" on page 7-14
NOTE
To obtain best accuracy, you should set the power levels prior to performing
the calibration. However, the analyzer compensates for nominal power
changes you make during a measurement, so that the error correction still
remains quite valid. In these cases, the Cor annunciator will change to C∆.
1-64
) will be much greater than the reverse isolation (S
21
.
COUPLED CH ON
Trans:FWD S21 (B/R)
Trans: REV S12 (A/R)
DUAL CHAN ON
1-51. If the port power levels are in different power ranges, one of
for information on how to override this state.
.
and set the power level for port
Power
and set the power level for port
Power
for error-correction procedures.
.
"Source Attenuator Switch
).
12
) and the
21
).
12