GPIB Operation

This section contains information on the following topics:
• local key
• GPIB controller modes
• instrument addresses
• using the parallel port

Key

Local
This key is allows you to return the analyzer to local (front panel) operation from remote
(computer controlled) operation. This key will also abort a test sequence or hardcopy
print/plot. In this local mode, with a controller still connected on GPIB, you can operate the
analyzer manually (locally) from the front panel. This is the only front panel key that is
not disabled when the analyzer is remotely controlled over GPIB by a computer. The
exception to this is when local lockout is in effect: this is a remote command that disables
the
key, making it difficult to interfere with the analyzer while it is under
Local
computer control.
In addition, the
Local
mode, and to the address menu, where you can enter the GPIB addresses of peripheral
devices and select plotter/printer ports. You can also set the mode of the parallel port here.
The GPIB menu consists of the following softkeys:
SYSTEM CONTROLLER
TALKER/LISTENER
USE PASS CONTROL
SET ADDRESS
PARALLEL [ ]
GPIB DIAG on OFF
DISK UNIT NUMBER
VOLUME NUMBER
The analyzer is factory-equipped with a remote programming interface using the General
Purpose Interface Bus (GPIB). This enables communication between the analyzer and a
controlling computer as well as other peripheral devices. This menu indicates the present
GPIB controller mode of the analyzer. Three GPIB modes are possible: system controller,
talker/listener, and pass control.
key accesses the GPIB menu, where you can set the controller
Operating Concepts
GPIB Operation
7-77