4. To define the THRU/LINE standard, press:
DEFINE STANDARD
SPECIFY OFFSET
STD DONE (DEFINED)
5. To define the LINE/MATCH standard, press:
DEFINE STANDARD
6. For the purposes of this example, change the name of the standard by pressing
MODIFY STD DEFINITION
and then modify the name to "MATCH".
7. When the title area shows the new label, press:
DONE STD DONE (DEFINED)
Assign the Standards to the Various LRM Classes
8. To assign the calibration standards to the various TRL calibration classes, press:
Cal
CAL KIT
9. Since you previously designated standard #1 for the REFLECT standard, press:
1
x1
10.Since you previously designated standard #3 for the LINE/MATCH standard, press:
TRL LINE OR MATCH
11.Since you previously designated standard #4 for the THRU/LINE standard, press:
4
TRL THRU
12.To complete the specification of class assignments, press:
SPECIFY CLASS DONE
Label the Classes
NOTE
To enter the following label titles, an external keyboard may be used for
convenience.
13.Press
LABEL CLASS
14.Change the label of the "TRL REFLECT" class to "LRMSHORT."
15.Change the label of the "TRL LINE OR MATCH" class to "LRMLOAD."
16.Change the label of the "TRL THRU" class to "LRMTHRU."
17.Press
LABEL CLASS DONE
4
x1
DELAY/THRU
OFFSET DELAY
3
x1
LOADS
LABEL STD
MODIFY
SPECIFY CLASS MORE
3
x1
x1
MORE
MORE
.
Calibrating for Increased Measurement Accuracy
MODIFY STD DEFINITION
0
x1
STD OFFSET DONE
ERASE TITLE
MORE
.
LRM Error Correction
, if a previous title exists,
TRL REFLECT
6-55