Agilent Technologies 8753ET User Manual page 169

Network analyzers
Hide thumbs Also See for 8753ET:
Table of Contents

Advertisement

Making Mixer Measurements
Isolation Example Measurements
8. Connect the external LO source to the mixer's LO port.
9. The measurement results show the mixer's RF feedthrough.
NOTE
You may see spurious responses on the analyzer trace due to interference
caused by LO to IF leakage in the mixer. This can be reduced with averaging
or by reducing the IF bandwidth.
Figure 2-38 Example Mixer RF Feedthrough Measurement
You can measure the IF to RF isolation in a similar manner, but with the following
modifications:
• Use the analyzer source as the IF signal drive.
• View the leakage signal at the RF port.
2-48