Perform the High Dynamic Range Measurement

1. Return the analyzer to the IF frequency range. Press
.
G/n
2. Make the connections shown in
3. Set the LO source to the desired CW frequency of 1500 MHz and power level to 13 dBm.
Figure 2-19 Connections for a High Dynamic Range Swept IF Conversion Loss
Measurement
4. Set the analyzers LO frequency to match the frequency of the LO source by pressing:
System
INSTRUMENT MODE
FREQUENCY: CW
5. To select the converter type and low-side LO measurement configuration, press:
RETURN
DOWN CONVERTER
In this low-side LO, down converter measurement, the analyzer's source frequency
range will be offset higher than the receiver frequency range. The source frequency
range can be determined from the following equation:
receiver frequency range (100 – 1000 MHz) + LO frequency (1500 MHz) = 1.6 − 2.5 GHz
6. To view the conversion loss in the best vertical resolution, press:
VIEW MEASURE
Figure
2-19.
FREQ OFFS MENU LO MENU
1500
M/µ
RF > LO FREQ OFFS ON
AUTOSCALE
Scale Ref
Making Mixer Measurements
High Dynamic Range Swept RF/IF Conversion Loss
Start
100
M/µ
Stop
1
2-23