Making Mixer Measurements
Conversion Loss Using the Frequency Offset Mode
Figure 2-11 Connections for R Channel and Source Calibration
6. From the front panel of the analyzer, set the desired receiver frequency and source
output power by pressing:
INSTRUMENT MODE
System
Start
100
Stop
350
If the LO frequency is not set to 0 Hz, press:
FREQ OFFS ON
Power
0
x1
7. To view the measurement trace, press:
INPUT PORTS
Meas
8. Select the analyzer as the system controller:
Local
SYSTEM CONTROLLER
9. Set the power meter's address:
SET ADDRESSES
ADDRESS: P MTR/HPIB
x1
10.Select the appropriate power meter by pressing
model number is displayed (HP/Agilent 436A or HP/Agilent 438A/437).
2-14
FREQ OFFS MENU
M/µ
M/µ
x1
R
(where aa is the GPIB address of the power meter)
aa
until the correct
POWER MTR [ ]