Figure 1-41 Duplexer Measurement
Normally, a 2-port calibration requires a forward and reverse sweep to complete before the
displayed trace updates. For faster tuning, it is possible to set the number of sweeps for the
active display channel (S
the inactive display channel. In this example we choose 8 updates of the forward
parameters to 1 update of the reverse in channel 1, and 8 updates of the reverse to 1
update of the forward in channel 2 (where the active parameters are S
Press
System CONFIGURE MENU TESTSET SW CONTINUOUS
Chan 1
Press
System CONFIGURE MENU TESTSET SW CONTINUOUS
Chan 2
and S
for channel 1 in this case) to update more often than
11
21
Making Measurements
Characterizing a Duplexer (ES Analyzers Only)
and S
22
).
12
.
8
x1
.
8
x1
1-53