Calibrating for Increased Measurement Accuracy
Calibration Considerations
Figure 6-1 Typical Responses of Calibration Standards after Calibration
CH1 S 11
1 U FS
Cor
1
START
.300
000 MHZ
7mm or Type-N Male
Short (No Offset)
CH1 S 11
1 U FS
Cor
START
.300
000 MHZ
7mm or Type-N Male
Open (No Offset) with Fringing Capacitance
Interpolated Error correction
You may want to use interpolated error correction when you choose a subset of a frequency
range that you already corrected, when you change the number of points, or when you
change to CW. This feature also allows you to change the parameters in a 2-port correction,
such as IF bandwidth, power, or sweep time. The analyzer calculates the systematic errors
from the errors of the original correction.
To activate interpolated measurement correction, press
ON is selected, and
CORRECTION on OFF
6-8
1: 998.84 mU
-179.92
3 000 000 000 MHz
.
STOP 3
000 000 000 MHZ
.
1: 998.25 mU
-11.779
3 000 000 000 MHz
.
1
STOP 3
000 000 000 MHZ
.
so that ON is selected. When interpolation is
CH1 S 11
1 U FS
1
Cor
START
.300
000 MHZ
Type-N Female,
3.5mm Male or Female Offset Short
CH1 S 11
1 U FS
Cor
START
.300
000 MHZ
Type-N Female
3.5 mm Male or Female Offset Open
INTERPOL ON off
Cal
1: 999.62 mU
142.07
3 000 000 000 MHz
.
STOP 3
000 000 000 MHZ
.
1: 999.1 mU
-44.561
3 000 000 000 MHz
.
1
STOP 3
000 000 000 MHZ
.
pa5162e
so that