Calibrating for Increased Measurement Accuracy
Calibration Considerations
• The standard is an open termination, which by definition exhibits a certain amount of
fringe capacitance (and therefore phase shift). Open terminations which are offset from
the mating plane will exhibit a phase shift due to the offset in addition to the phase
shift caused by the fringe capacitance.
The most important point to remember is that these properties will not affect your
measurements. The analyzer compensates for them during measurement. As a result, if
these standards are measured after a calibration, they will not appear to be "perfect"
shorts or opens. This is an indication that your analyzer is working properly and that it has
successfully performed a calibration.
calibration standards after calibration.
Electrical Offset
Some standards have reference planes that are electrically offset from the mating plane of
the test port. These devices will show a phase shift with respect to frequency.
shows which reference devices exhibit an electrical offset phase shift. The amount of phase
shift can be calculated with the formula:
Φ = (360° × f × l)/c where:
f = frequency
l = electrical length of the offset
c = speed of light (3 x 10
Fringe Capacitance
All open circuit terminations exhibit a phase shift over frequency due to fringe
capacitance. Offset open circuits have increased phase shift because the offset acts as a
small length of transmission line. Refer to
Table 6-1 Calibration Standard Types and Expected Phase Shift
Test Port Connector Type
7-mm
Type-N male
3.5-mm male
3.5-mm female
2.4-mm male
2.4-mm female
Type-N female
75Ω Type-N female
7-mm
Type N-male
6-6
Figure 6-1
8
meters/second)
Table
Standard Type
Short
Offset Short
Open
shows sample displays of various
6-1.
Expected Phase Shift
180°
(
360°
180°
+
--------------------------------- -
Φ°
+
capitance
Table 6-1
×
×
)
f
l
c

Advertisement