Operating Concepts
Measurement Calibration
Enhanced Response Calibration Error Model
Enhanced response calibration uses the same error model as the forward configuration
portion of
Figure
7-38. In the response portion, the source and load match effects are
fully-accounted for giving the same accuracy to the forward tracking term (E
) as the
TF
two-port calibration. During the measurement, the enhanced response calibration
performs a correction which is mathematically the same as setting the values of E
, E
,
LF
LR
E
, E
, E
to zero (0) and the values of E
and E
to one (1) in the equations for S11
SR
XR
DR
RR
TR
and S21 shown in
Figure 7-39
Figure 7-39 Full Two-Port Error Model Equations