Calibrating for Increased Measurement Accuracy
Full Two-Port Error Correction (ES Analyzers Only)
17.To compute the error coefficients, press:
DONE 2-PORT CAL
The analyzer displays the corrected measurement trace. The analyzer also shows the
notation Cor at the left of the screen, indicating that error correction is on.
You can save or store the measurement correction to use for later
NOTE
measurements. Refer to
Measurement Results"
18.This completes the full two-port correction procedure. You can connect and measure
your device under test.
6-32
Chapter 4 , "Printing, Plotting, and Saving
for procedures.