Optimizing Measurement Results
Reducing Receiver Crosstalk
Reducing Receiver Crosstalk
To reduce receiver crosstalk you can do the following:
• Perform a response and isolation measurement calibration.
• Set the sweep to the alternate mode.
Alternate sweep is intended for measuring wide dynamic range devices, such as high pass
and bandpass filters. This sweep mode removes a type of leakage term through the device
under test, from one channel to another.
To set the alternate sweep, press
.
MORE
ALTERNATE A and B
Cal
Refer to
"Frequency Response and Isolation Error Corrections" on page
6-17.
5-18