2400 Series SourceMeter
User's Manual
Source-measure programming example
Table 3-6
summarizes the command sequence for a basic source-measure procedure. Note
that the steps correspond to those listed previously in
dure." These commands set up the SourceMeter as follows:
•
Source function and range: volts, 20V
•
Source mode: fixed
•
Source output level: 10V
•
Current compliance: 10mA
•
Measure function and range: current, 10mA
Table 3-6
Basic source-measure programming example
1
Step
Action
1
Select source function, mode
2
Set source range, level, compliance
3
Set measure function, range
4
Turn on output
5
Read data
6
Turn off output
1. Steps correspond to front panel steps listed previously in
2. Commands must be sent in order given.
3. Instrument must be addressed to talk after :READ? to acquire data.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Basic Source-Measure Operation
"Front panel source-measure proce-
2,3
Commands
*RST
Restore GPIB defaults.
:SOUR:FUNC VOLT
Select voltage source.
:SOUR:VOLT:MODE FIXED
Fixed voltage source mode.
:SOUR:VOLT:RANG 20
Select 20V source range.
:SOUR:VOLT:LEV 10
Source output = 10V.
:SENS:CURR:PROT 10E-3
10mA compliance.
:SENS:FUNC "CURR"
Current measure function.
:SENS:CURR:RANG 10E-3
10mA measure range.
:FORM:ELEM CURR
Current reading only.
:OUTP ON
Output on before measuring.
:READ?
Trigger, acquire reading.
:OUTP OFF
"Front panel source-measure
procedure."
3-19
Comments