Questionable Event Register - Keithley SourceMeter 2400 User Manual

Sourcemeter 2400 series source measure unit
Hide thumbs Also See for SourceMeter 2400:
Table of Contents

Advertisement

®
2400 Series SourceMeter
User's Manual

Questionable Event Register

The used bits of the Questionable Event Register (shown in
follows:
Bits B0 through B7 — Not used.
Bit B8, Calibration Summary (Cal) — Set bit indicates that an invalid calibration
constant was detected during the power-up sequence. This error will clear after
successful calibration of the instrument.
Bits B9 through B13 — Not used.
Bit B14, Command Warning (Warn) — Set bit indicates that a Signal Oriented
Measurement Command parameter has been ignored.
Bit B15 — Not used.
Figure 15-7
Questionable event status
:stat:ques:cond?
:stat:ques?
OR
To QSB of Status
Byte Register.
(See Figure 15-3.)
:stat:ques:enab<NRf>
stat:ques:enab?
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Warn
CAL
(B15)
(B14)
(B13 - B9)
(B8)
Warn
CAL
(B15)
(B14)
(B13 - B9)
(B8)
&
&
Warn
CAL
(B15)
(B14)
(B13 - B9)
(B8)
Warn = Command Warning
& = Logical AND
Cal = Calibration Summary
OR = Logical OR
Status Structure
15-15
Figure
15-7) are described as
Questionable
Condition
(B7 - B0)
Register
Questionable
Event Register
(B7 - B0)
Questionable
Event Enable
(B7 - B0)
Register