Guard Sense - Keithley SourceMeter 2400 User Manual

Sourcemeter 2400 series source measure unit
Hide thumbs Also See for SourceMeter 2400:
Table of Contents

Advertisement

6-38
Source-Measure Concepts

Guard sense

When the GUARD-to-LO resistance path is less than 1kΩ, remote guard sensing should be
used to compensate for IR drop in the GUARD test lead and/or switch contacts on a
switching card.
Figure 6-19
was modified to create
showing the 1Ω resistance (R
GUARD-to-LO is less than 1kΩ, the IR drop in the guard test lead (R
cant. The guard voltage applied to the bottom of R2 is now significantly lower than the In/
Out HI voltage of the SourceMeter. As a result, leakage current (I
adversely affecting the resistance measurement of R1.
The guard test lead IR drop is compensated for by connecting GUARD SENSE as shown
in
Figure
6-20B. Sensing allows the guard voltage to be sensed (measured) at the resistor
network for better guard voltage regulation. If the remotely sensed guard voltage is less
than the output voltage of the SourceMeter, the guard voltage will be increased until the
sensed guard voltage equals the output HI voltage.
Note that in order to ensure that guard current (I
output voltage from the SourceMeter must not exceed 5V (50mA × 100Ω = 5V).
NOTE
Guard sense operation is automatic. There is no menu selection to enable or dis-
able guard sense.
For 6-wire ohms guard measurements, use the GUARD output-off state. For
details on the GUARD output-off state, see
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
2400 Series SourceMeter
Figure 6-20A
by changing the value of R3 to 100Ω and
) of the GUARD test lead. Since the resistance path from
TL
L
) in
Figure 6-20
does not exceed 50mA,
G
Section
13,
"Output-off
®
User's Manual
) becomes signifi-
TL
) flows through R2,
states."

Advertisement

Table of Contents
loading

Table of Contents