*SAV, *RCL programming example

Table 16-3
summarizes the basic command sequence for saving and recalling a setup. The
present setup is stored in memory location 2, GPIB defaults are restored, and the memory
location 2 setup is recalled.
Table 16-3
*SAV, *RCL programming example
Command
*SAV 2
Save present setup in memory location 2.
*RST
Restore GPIB defaults.
*RCL 2
Recall location 2 setup.
*RST — reset
When the *RST command is sent, the SourceMeter performs the following operations:
Returns the SourceMeter to the GPIB default conditions (see "Default" column of
SCPI tables in
Cancels all pending commands.
Cancels response to any previously received *OPC and *OPC? commands.
*TRG — trigger
Use the *TRG command to issue a GPIB trigger to the SourceMeter. It has the same effect
as a group execute trigger (GET).
Use the *TRG command as an event to control operation. The SourceMeter reacts to this
trigger if BUS is the programmed arm control source. The control source is programmed
from the TRIGger subsystem.