Rsynhronize <B> - Keithley SourceMeter 2400 User Manual

Sourcemeter 2400 series source measure unit
Hide thumbs Also See for SourceMeter 2400:
Table of Contents

Advertisement

®
2400 Series SourceMeter
User's Manual
2425 and 2430 DC Mode
<n> =
2430 Pulse Mode
<n> =
2440
<n> =
Query
:LEVel?
:LEVel? DEFault
:LEVel? MINimum
:LEVel? MAXimum
Description
This command is used to set compliance limits. A current compliance
limit is set for the V-Source, and a voltage compliance limit is set for the
I-Source. The SourceMeter cannot source levels that exceed these speci-
fied limits.
The :SENSe:CURRent:PROTection[:LIMit] command is used to set the
current compliance for the V-Source and the :SENSe:VOLTage:
PROTection[:LIMit] command is used to set the voltage compliance for
the I-Source.
NOTE
You cannot set compliance less than 0.1% of the present measurement range.
RSYNhronize <b>
[:SENSe[1]]:CURRent[:DC]:PROTection:RSYNchronize <b>
[:SENSe[1]]:VOLTage[:DC]:PROTection:RSYNchronize <b>
Parameters
<b> =
Description
Turning this feature on will make the measurement range follow the
compliance range setting when autoranging is off. When the compliance
value is set, the measurement range will be on the same range as the
compliance setting. The two commands above change one setting;
changing the state of one will automatically change the other.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
SCPI Command Reference
-3.15 to 3.15
Current compliance limit
-105 to 105
Voltage compliance limit
DEFault
105uA, 21V
MINimum
-3.15A, -105V
MAXimum
3.15A, 105V
-10.5 to 10.5
Current compliance limit
-105 to 105
Voltage compliance limit
DEFault
105uA, 21V
MINimum
-10.5A, -105V
MAXimum
10.5A, 105V
-5.25 to 5.25
Current compliance limit
-42 to 42
Voltage compliance limit
DEFault
105uA, 21V
MINimum
-5.25A, -42V
MAXimum
5.25A, 42V
Query compliance value
Query *RST default compliance
Query minimum allowable compliance
Query maximum allowable compliance
Control current measure and compliance
range sync
Control voltage measure and compliance
range sync
OFF
Disable range sync
ON
Enable range sync
18-69