Figure 13-2 Sink Operation; Sot Line - Keithley SourceMeter 2400 User Manual

Sourcemeter 2400 series source measure unit
Hide thumbs Also See for SourceMeter 2400:
Table of Contents

Advertisement

®
2400 Series SourceMeter
User's Manual Digital I/O Port, Safety Interlock, and Output Configuration

SOT line

The input line (SOT) is used by the handler to start limit testing. With the
event selected
(Section
order to provide event detection which starts the testing process. With the
event selected, the handler must pulse SOT high in order to provide event detection, which
starts the testing process. With
the testing process.
Interlock line
The interlock line is intended for use with a safety interlock on a device handler or test fix-
ture. See
"Safety interlock," page
+5V output
The Digital I/O Port provides a +5V output that can be used to drive external logic cir-
cuitry. Maximum current output for this line is 300mA. This line is protected by a self-
resetting fuse (one hour recovery time).
Digital output configuration
There are two basic methods to connect external components to the digital output lines,
sink operation and source operation.
Sink operation
Figure 13-2
shows the basic output configuration for sink operation. Note that the external
relay coil is connected between the digital output line (pins 1 to 4) and +5V (pin 7). With
this configuration, the digital output line must be set LO to energize the relay, and the
maximum sink current is 500mA.
Figure 13-2
Sink operation
SourceMeter
+5V
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
11,
"Configuring
triggering"), the handler must pulse SOT low in
↑↓
STEST selected, either a high or low SOT pulse starts
13-5, for more details.
External
Pin 7
Relay
Pins 1-4
Maximum sink current: 500mA
Pin 9
Digital I/O
Port
13-3
STEST arm
STEST arm
To other
Circuits