Typical Nplc Cache Test Times - Keithley SourceMeter 2400 User Manual

Sourcemeter 2400 series source measure unit
Hide thumbs Also See for SourceMeter 2400:
Table of Contents

Advertisement

®
2400 Series SourceMeter
User's Manual

Typical NPLC cache test times

Typically, NPLC caching will decrease source memory sweep times by a factor of three.
The table below shows typical averaged times for a test consisting of 10 sweeps of four
source memory locations with NPLC values for successive memory locations set to 10, 1,
0.1, and 0.01 respectively.
NPLC cache conditions
Auto Zero OFF, Caching OFF
Auto Zero ON, Caching OFF
Auto Zero OFF, Caching ON
Auto Zero ON, Caching ON
Auto Zero OFF, Caching ON, Cache empty
V-source protection
Use V-source protection to select the maximum voltage level the SourceMeter can output.
Available settings are listed as follows:
SourceMeter
2400
20V, 40V, 60V, 80V, 100V, 120V, 160V, NONE (>160V)
2410
20V, 40V, 100V, 200V, 300V, 400V, 500V, NONE (>500V)
2420
6V, 12V, 18V, 24V, 30V, 36V, 48V, NONE (>48V)
2425/2430
10V, 20V, 30V, 40V, 50V, 60V, 80V, NONE (>80V)
2440
4V, 8V, 12V, 16V, 20V, 24V, 32V, NONE (>40V)
These are absolute values with 5% tolerance. The power-on default is NONE.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Basic Source-Measure Operation
Time
5.89s
5.89s
2.05s
5.89s
2.5s
V-Source protection limit settings
3-11