Changing MATH function

If you started with only a basic measurement function selected, you can enable a
MATH function, but only the voltage, current, or resistance component of the cal-
culation will be stored in the buffer. The results of the MATH function will not be
stored.
If you started with a MATH function enabled, you can select a different MATH
function. The results of the new MATH function are stored in the buffer.
If you started with REL and/or LIMITS enabled, the buffer will stop storing read-
ings if you select a MATH function. Storage will continue if you return to the orig-
inal configuration.
See
Section 8
for more information on MATH.
Changing REL or LIMITS
If you started with only a basic measurement function selected, you can enable
REL and/or LIMITS, but only the voltage, current, or resistance component of the
operation will be stored in the buffer. The results of REL and/or LIMITS are not
stored.
If you started with a MATH function enabled, only the result of the MATH calcula-
tion will be stored in the buffer if REL and/or LIMITS is enabled.
If you started with REL and/or LIMITS enabled, you can change REL and or
LIMITS. The results of new REL and/or LIMITS are stored in the buffer.