Triggered[:Amplitude] <N> - Keithley SourceMeter 2400 User Manual

Sourcemeter 2400 series source measure unit
Hide thumbs Also See for SourceMeter 2400:
Table of Contents

Advertisement

®
2400 Series SourceMeter
User's Manual
The MINimum and MAXimum parameters are only valid if the highest
source range is presently selected. Sending the MINimum or MAXi-
mum parameters on a lower source range will generate error -221 (Set-
ting Conflict).
TRIGgered[:AMPLitude] <n>
:SOURce[1]:CURRent[:LEVel]:TRIGgered[:AMPLitude] <n>
:SOURce[1]:VOLTage[:LEVel]:TRIGgered[:AMPLitude] <n>
Parameters
2400
<n> =
2410
<n> =
2420
<n> =
2425 and 2430 DC Mode
<n> =
2430 Pulse Mode
<n> =
2440
<n> =
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
SCPI Command Reference
Set fixed I-Source amplitude when
triggered
Set fixed V-Source amplitude when
triggered
-1.05 to 1.05
Set I-Source amplitude (amps)
-210 to 210
Set V-Source amplitude (volts)
DEFault
0A or 0V
MINimum
-1.05A or -210V
MAXimum
+1.05A or +210V
-1.05 to 1.05
Set I-Source amplitude (amps)
-1100 to 1100
Set V-Source amplitude (volts)
DEFault
0A or 0V
MINimum
-1.05A or -1100V
MAXimum
+1.05A or +1100V
-3.15 to 3.15
Set I-Source amplitude (amps)
-63 to 63
Set V-Source amplitude (volts)
DEFault
0A or 0V
MINimum
-3.15A or -63V
MAXimum
+3.15A or +63V
-3.15 to 3.15
Set I-Source amplitude (amps)
-105 to 105
Set V-Source amplitude (volts)
DEFault
0A or 0V
MINimum
-3.15A or -105V
MAXimum
+3.15A or +105V
-10.5 to 10.5
Set I-Source amplitude (amps)
-105 to 105
Set V-Source amplitude (volts)
DEFault
0A or 0V
MINimum
-10.5A or -105V
MAXimum
+10.5A or +105V
-5.25 to 5.25
Set I-Source amplitude (amps)
-42 to 42
Set V-Source amplitude (volts)
DEFault
0A or 0V
MINimum
-5.25A or -42V
MAXimum
+5.25A or +42V
18-79

Advertisement

Table of Contents
loading

Table of Contents