Status Subsystem - Keithley SourceMeter 2400 User Manual

Sourcemeter 2400 series source measure unit
Hide thumbs Also See for SourceMeter 2400:
Table of Contents

Advertisement

18-104 SCPI Command Reference

STATus subsystem

The STATus subsystem is used to control the status registers of the SourceMeter. The com-
mands in this subsystem are summarized in
NOTE
These registers and the overall status structure are fully explained in
"Status
Structure."
Read event registers
[:EVENt]?
:STATus:MEASurement[:EVENt]?
:STATus:QUEStionable[:EVENt]?
:STATus:OPERation[:EVENt]?
Description
These query commands are used to read the contents of the status event
registers. After sending one of these commands and addressing the
SourceMeter to talk, a value is sent to the computer. This value indicates
which bits in the appropriate register are set.
Program event enable registers
ENABle <NDN> or <NRf>
:STATus:MEASurement:ENABle <NDN> or <NRf>
:STATus:QUEStionable:ENABle <NDN> or <NRf>
:STATus:OPERation:ENABle <NDN> or <NRf>
Description
These commands are used to set the contents of the event enable
registers.
Read condition registers
CONDition?
:STATus:MEASurement:CONDition?
:STATus:QUEStionable:CONDition?
:STATus:OPERation:CONDition?
Description
These query commands are used to read the contents of the condition
registers.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
®
2400 Series SourceMeter
User's Manual
Table
18-8.
Section
Read Measurement Event Register
Read Questionable Event Register
Read Operation Event Register
Program Measurement Event Enable
Register
Program Questionable Event Enable
Register
Program Operation Event Enable Register
Read Measurement Condition
Read Questionable Register
Read Operation Condition
15,