Event enable registers
As
Figure 15-1
shows, each status register set has an enable register. Each event register
bit is logically ANDed (&) to a corresponding enable bit of an enable register. Therefore,
when an event bit is set and the corresponding enable bit is set (as programmed by the
user), the output (summary) of the register will set to 1, which in turn sets the summary bit
of the Status Byte Register.
The commands to program and read the event enable registers are listed in
details on programming and reading registers, see
15-5, and
"Reading registers," page
NOTE
The bits of any enable register can be reset to 0 by sending the 0
parameter value with the appropriate enable command (i.e.
STATus:OPERation:ENABle 0).
Table 15-7

Event enable registers commands

Command
*ESE <NDN> or <NRf>
Program Standard Event Enable Register. (See
*ESE?
Read Standard Event Enable Register.
STATus
STATus Subsystem:
:OPERation
:ENABle <NDN> or <NRf>
:ENABle?
:MEASurement
:ENABle <NDN> or <NRf>
:ENABle?
:QUEStionable
:ENABle <NDN> or <NRf>
:ENABle?
Parameters:
<NDN> = #Bxx...x Binary format (each x = 1 or 0)
= #Hx Hexadecimal format (x = 0 to FFFF)
= #Qx Octal format (x = 0 to 177777)
<NRf> = 0 to 65535 Decimal format
Note: Power-up and STATus:PRESet reset all bits of all enable registers to 0. *CLS has no effect.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
"Programming enable registers," page
15-6.
Description
"Parameters.")
Operation Event Enable Register:
Program enable register. (See "Parameters.")
Read enable register.
Measurement Event Enable Register:
Program enable register. (See "Parameters.")
Read enable register.
Questionable Event Enable Register:
Program enable register. (See "Parameters.")
Read Measurement Event Enable Register:
Status Structure