Performing Front Panel Limit Tests - Keithley SourceMeter 2400 User Manual

Sourcemeter 2400 series source measure unit
Hide thumbs Also See for SourceMeter 2400:
Table of Contents

Advertisement

®
2400 Series SourceMeter
User's Manual
Figure 12-10
Limits configuration menu tree
H/W
DIGOUT
LIMITS
AUTO
MODE
SIZE
CONTROL
CLEAR

Performing front panel limit tests

Perform the basic steps below to run limit tests from the front panel. See
testing," page
12-23, for remote commands and a programming example.
Step 1: Configure test system.
As previously explained, your test system could be as simple as connecting a DUT to the
SourceMeter
(Section
operations. Adding a scanner to the test system allows you to test multi-element devices
(such as resistor networks).
Step 2: Configure source-measure functions.
Configure the SourceMeter for the desired source-measure operations as follows:
1.
Select the desired source function by pressing SOURCE V or SOURCE I.
2.
Set the source level and compliance limit to the desired values.
3.
Press MEAS V or MEAS I to select the desired measurement function, then choose
the desired measurement range.
Refer to
Section
3,
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
CONFIG
LIMIT
S/W
PASS
LIMITS
DIG I/O
FAIL
LIM2,3
MODE
LIM5-12
PATTERN
CONTROL
LOLIM
HILIM
2, "Connections") or could employ the use of a handler for binning
"Basic source-measure
procedure," for more information.