Introduction

This Appendix contains a summary of status and error messages, which status register bits
are set when messages occur, and methods to avoid or eliminate most common SCPI
errors.
Status and error messages
Table B-1
summarizes status and error messages, which are stored in the Error Queue.
Each message is preceded by a code number. Negative (-) numbers are used for SCPI
defined messages, and positive (+) numbers are used for Keithley defined messages. Note
that error and status conditions will also set specific bits in various status registers, as sum-
marized in
Table
B-1.
Section 15
has detailed information on registers and queues. Briefly, you can use the fol-
lowing queries to obtain error and status information:
:SYST:ERR? — reads Error Queue.
*ESR? — reads Standard Event Status Register.
:STAT:OPER? — reads Operation Event Register.
:STAT:MEAS? — reads Measurement Event Register.
:STAT:QUES? — reads Questionable Event Register.
NOTE
SCPI confirmed messages are described in volume 2: Command Reference of
the Standard Commands for Programmable Instruments. Refer to the
:SYSTem:ERRor? command.