Disable/Enable Auto Zero - Keithley SourceMeter 2400 User Manual

Sourcemeter 2400 series source measure unit
Hide thumbs Also See for SourceMeter 2400:
Table of Contents

Advertisement

®
2400 Series SourceMeter
User's Manual
4.
For continuous pulse output, select INFINITE, press ENTER and proceed to
step 6. Otherwise, select FINITE, press ENTER and proceed to the next step to set
the arm count.
5.
Enter the desired arm count value and press ENTER.
6.
Use the EXIT key to back out of the menu.
Perform the following steps to set the trigger count:
1.
Press CONFIG and then TRIG to display the trigger configuration menu.
2.
Select TRIG LAYER from the displayed choices and press ENTER.
3.
Select COUNT from the displayed choices and press ENTER.
4.
Enter the desired trigger count value and press ENTER.
5.
Use the EXIT key to back out of the menu.

Disable/enable auto zero

Pulse speed can be increased by disabling auto zero. However, zero drift will eventually
corrupt the accuracy of the pulse measurement. To re-establish an accurate pulse measure-
ment, enable auto zero or change the speed setting.
1.
Press MENU to display the main menu.
2.
Select A/D CTRL from the displayed choices and press ENTER.
3.
Select AUTO ZERO from the displayed choices and press ENTER.
4.
Select DISABLE or ENABLE and press ENTER.
5.
Use the EXIT key to back out of the menu.
Remote command Pulse Mode configuration
For remote operation, the commands to select and configure the Pulse Mode are provided
in
Table
5-1.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Pulse Mode Operation (Model 2430 only)
5-11