Bench Defaults - Keithley SourceMeter 2400 User Manual

Sourcemeter 2400 series source measure unit
Hide thumbs Also See for SourceMeter 2400:
Table of Contents

Advertisement

®
2400 Series SourceMeter
User's Manual

Bench defaults

The bench defaults are listed as follows. They are also denoted in
symbol.
Arm-In Event = Immediate
Trigger-In Source = Immediate
Arm Count = 1
Trigger Count = 1
Trigger Delay = 0.0 sec
Delay Action = 0.001 sec
Source Trigger In Event = On
Delay Trigger In Event = Off
Measure Trigger In Event = Off
Trigger Out Events = All Trigger Out Events are disabled (off)
Arm Out Event = Off
Event Detection Bypasses = Never (both layers)
When the output is turned ON, the SourceMeter will run in a continuous loop around the
trigger model. After each Measure Action, operation will continue at the top of the trigger
model. The SourceMeter can be returned to idle by turning the output OFF.
Operation summary
The trigger model is designed to offer versatility for the various source-measure applica-
tions. Typically, it allows you to perform a specified number of measurements at various
source levels.
For example, assume you want to perform three measurements each at two different
V-source levels (1V and 2V). To do this, set the arm count to two, the trigger count to
three, and use a 6-point Custom Sweep configured as follows:
P0000 = 1V
P0001 = 1V
P0002 = 1V
P0003 = 2V
P0004 = 2V
P0005 = 2V
When the sweep is started, operation falls into the Trigger Layer and performs three mea-
surements at the 1V source level. Operation then loops back into the Trigger Layer to per-
form three measurements at the 2V source level. The six readings are stored in the buffer.
Note that after the sweep is finished, the SourceMeter does not return to idle. Operation
continues at the top of the trigger model. Subsequent measurements are performed at the
2V level and are not stored in the buffer.
For details on the Custom Sweep, see
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Triggering
Figure 11-1
Section
10,
"Sweep
Operation."
11-7
by the " "

Advertisement

Table of Contents
loading

Table of Contents