Signal Oriented Measurement Command Summary; Figure - Keithley SourceMeter 2400 User Manual

Sourcemeter 2400 series source measure unit
Hide thumbs Also See for SourceMeter 2400:
Table of Contents

Advertisement

17-2
SCPI Signal Oriented Measurement Commands
Command summary
The signal oriented measurement commands are used to acquire readings. You can use
these high-level instructions to control the measurement process. These commands are
summarized in
Table
Table 17-1

Signal oriented measurement command summary

Command
:CONFigure:<function>
:CONFigure?
:FETCh?
:READ?
:MEASure[:<function>]?
Configuring measurement function
NOTE
For the Model 2430, the :CONFigure commands are not valid while in the Pulse
Mode. They result in error +831: Invalid in Pulse Mode.
:CONFigure:<function>
Parameters
<function> = CURRent[:DC]
Query
:CONFigure?
Description
This command configures the instrument to a specific setup for mea-
surements on the specified function. The :READ? command is then typ-
ically used to trigger a specified number of measurements. See
":READ?."
When this command is sent, the SourceMeter will be configured as
follows:
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
2400 Series SourceMeter
17-1.
Description
Configures SourceMeter for measurements on specified
function. Output turns on.
Returns active function(s).
Requests latest readings.
Performs an :INITiate and a :FETCh?.
One-shot measurement mode. Performs a :CONFigure and
a :READ?.
Amps function
VOLTage[:DC]
Volts function
RESistance
Ohms function
Returns active function(s).
Select specified function.
All controls related to the selected function are defaulted to the
*RST values.
The event control source of the Trigger Model is set to Immediate.
®