Pulse Mode Operation (Model 2430 only)

Step 7: Turn output on.

Turn the output on by pressing the TRIG key or the ON/OFF OUTPUT key. The OUTPUT
indicator will be on whenever a pulse is being sourced. During each pulse off-time, the
output will be off (indicator off).
If configured to output a finite number of pulses, the output will disable after the last pulse
is sourced. While sourcing pulses, you can disable the output at any time by pressing the
ON/OFF OUTPUT key.
Pulse-ohms measurements
NOTE
Ohms source readback cannot be used for the Pulse Mode. Make sure that the
source is not in compliance when taking pulse-ohms readings.
"Ohms source
Auto and Manual pulse-ohms measurements
While in the Pulse Mode, the Model 2430 can perform auto or manual ohms measure-
ments. After selecting and configuring the Pulse Mode as explained in this section, per-
form the
"Auto ohms
procedure in
Section
For remote operation, see
basic ohms measurements.
Offset-compensated pulse-ohms measurements
While in the Pulse Mode, offset-compensated ohms is available as a math function
(FCTN). After selecting and configuring the Pulse Mode as explained in this section, refer
to
Section
8,
"Math
For remote operation, see
form offset-compensated ohms.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
2400 Series SourceMeter
readback."
measurements" procedure or the
"Manual ohms
4.
Section
4,
"Remote ohms
programming," for the commands for
operations," to perform offset-compensated ohms measurements.
Section
8,
"Remote math
operations," for the commands to per-
®
User's Manual
Section 4
covers
measurements"

Advertisement

Table of Contents
loading

Table of Contents