Remote Commands For Basic Ohms Measurements; Table; Remote Ohms Programming - Keithley SourceMeter 2400 User Manual

Sourcemeter 2400 series source measure unit
Hide thumbs Also See for SourceMeter 2400:
Table of Contents

Advertisement

4-12
Ohms Measurements
6.
Press EXIT to return to normal display.
Press MEAS then Ω to select the ohms measurement function.
7.
8.
Select the appropriate measurement range, or use autoranging if desired.
9.
Turn on the output by pressing the ON/OFF OUTPUT key.
10.
Take readings from the display.
11.
Turn the output off when done by pressing the ON/OFF OUTPUT key.

Remote ohms programming

The following paragraphs summarize those basic commands necessary for remote ohms
programming and also give a programming example for a typical ohms measurement
situation.
Remote ohms commands

Table 4-2

summarizes the remote commands for making basic ohms measurements. See
Section 18
for more details on these commands.
Table 4-2

Remote commands for basic ohms measurements

Command
:SENSe:FUNCtion "RESistance"
:SENSe:RESistance:RANGe <n>
:SENSe:RESistance:MODE <name>
:SENSe:RESistance:OCOMpensated <state>
:SENSe:VOLTage:PROTection <n>
:SENSe:CURRent:PROTection <n>
:SOURce:FUNCtion <name>
:SYSTem:RSENse <state>
:OUTPut <state>
:READ?
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
2400 Series SourceMeter
Description
Select ohms function.
Select ohms range (n = range).
Select ohms mode (name = MANual or AUTO).
Enable/disable offset compensation (state = ON or OFF).
Set voltage compliance n for manual ohms.
Set current compliance n for manual ohms.
Select source function (name = VOLTage or CURRent).
2-wire/4-wire sensing (state = ON, 4-wire; OFF, 2-wire).
Turn output on or off (state = ON or OFF).
Trigger and acquire reading.
®
User's Manual