2400 Series SourceMeter
User's Manual
DIRection <name>
:SOURce[1]:SWEep:DIRection <name>
Parameters
<name> =
Query
:DIRection?
Description
Normally, a sweep is run from the start level to the stop level. The
:STARt and :STOP, or :CENTer and :SPAN commands are used to set
these levels.
This command lets you change the execution direction of the sweep.
With DOWn selected, the sweep will begin at the stop level and end at
the start level. Selecting UP restores sweep operation to the normal start
to stop direction.
Configure list
CURRent <NRf list>
VOLTage <NRf list>
:SOURce[1]:LIST:CURRent <NRf list>
:SOURce[1]:LIST:VOLTage <NRf list>
Parameters
<NRf list> = NRf, NRf ... NRf
2400
NRf =
2410
NRf =
2420
NRf =
2425 and 2430 DC Mode
NRf =
2430 Pulse Mode
NRf =
2440
NRf =
Query
:CURRent?
:VOLTage?
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
SCPI Command Reference
Set direction of sweep
UP
Run sweep from start to stop
DOWn
Run sweep from stop to start
Query direction of sweep
Define I-Source list
Define V-Source list
-1.05 to 1.05
I-Source value
-210 to 210
V-Source value
-1.05 to 1.05
I-Source value
-1100 to 1100
V-Source value
-3.15 to 3.15
I-Source value
-63 to 63
V-Source value
-3.15 to 3.15
I-Source value
-105 to 105
V-Source value
-10.5 to 10.5
I-Source value
-105 to 105
V-Source value
-5.25 to 5.25
I-Source value
-42 to 42
V-Source value
Query I-Source list
Query V-Source list