RECall <NRf>
:SOURCe:MEMory:RECall <NRf>
Parameters
<NRf> =
Description
This command is used to return the SourceMeter to the setup stored at
the specified memory location.
Set scaling factor
TRIGgered:SFACtor <n>
:SOURce[1]:CURRent[:LEVel]:TRIGgered:SFACtor <n>
:SOURce[1]:VOLTage[:LEVel]:TRIGgered:SFACtor <n>
Parameters
<n> = -999.9999e+18 to 999.9999e+18
Query
:SFACtor?
Description
:SFAC instructs the SourceMeter to source the scaling factor times the
previous source memory location value. For example, if 10.0V is stored
in the first source memory (Source I, Measure V Mode), and the unit is
in the Source V, Measure I mode with :SFAC set to 0.1 and enabled, the
unit will output 1.0V for the second source memory location.
NOTE
These commands work only with source memory sweeps.
TRIGgered:SFACtor:STATe <b>
:SOURce[1]:CURRent[:LEVel]:TRIGgered:SFACtor:STATe <b> Enable/disable current scaling
:SOURce[1]:VOLTage[:LEVel]:TRIGgered:SFACtor:STATe <b> Enable/disable voltage scaling
Parameters
<b> =
Query
:SFACtor:STATe?
Description
:SFAC:STAT enables or disables scaling.
NOTE
These commands work only with source memory sweeps.
Sweep and list program examples
Linear voltage sweep
Linear voltage sweep from 1V to 10V in 1V increments:
*RST
SOUR:FUNC:MODE VOLT
SOUR:SWE:SPAC LIN
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
2400 Series SourceMeter
Return to specified setup
1 to 100
Specify memory location
Set current scaling factor
Set voltage scaling factor
Scaling factor
Query scaling factor
1 or ON
Enable scaling
0 or OFF
Disable scaling
Query enabled/disabled scaling state
®