Figure 11-1 Front Panel Trigger Model - Keithley SourceMeter 2400 User Manual

Sourcemeter 2400 series source measure unit
Hide thumbs Also See for SourceMeter 2400:
Table of Contents

Advertisement

®
2400 Series SourceMeter
User's Manual
Figure 11-1
Front panel trigger model
Never
Immediate
GPIB
Timer
Manual
Arm-In
TLink
Event
Stest
Stest
↑↓
Stest
Arm-Out Event
Never
On/Off
Trigger Out Event
On/Off
Immediate
Trigger-In
Trigger Link
Source
Trigger Out Event
On/Off
= Bench Default
= Output Trigger
Trigger Out Event
Soak time takes the place of the delay time only during the first SDM cycle after initial sweep
*
trigger if the unit is in the MULTIPLE mode. See Section 7, "Autorange change mode."
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Idle
Turn Output ON
Once
Bypass
No
Arm Event
Detector
?
Another
Yes
Arm
?
Arm Event
Detector
Arm-Out Event
On/Off
On/Off
Once
Bypass
Source Event
No
Detector
?
Another
Yes
Trigger
?
Source Event
Detector
Trigger Delay
0.0 sec
SOURCE
Action
On/Off
Delay Event
Detector
DELAY
0.001 sec
Action*
On/Off
Measure
Event
Detector
MEASURE
Action
On/Off
Triggering
11-3
Idle
Arm
Arm
Counter
Layer
1
Trigger
Layer
Trigger
Counter
1
MEASURE
Action
CONV
CONV
Filter
Process
(Repeat)
CONV
CONV = Reading Conversion

Advertisement

Table of Contents
loading

Table of Contents