5-16
Pulse Mode Operation (Model 2430 only)

Pulse-measure programming example

Table 5-2

summarizes the command sequence to output and measure pulses. Note that the
steps correspond to those listed in the
Table 5-2

Basic pulse programming example

1
Step
Action
1
Select Pulse Mode, and set pulse
width and pulse delay.
2
Set pulse measurement speed.
3
Set pulse count.
4
Select source.
5
Set source range, level, compliance.
6
Set measure function, range.
7
Turn on output and read data.
1. Steps correspond to front panel steps in
"Front panel pulse-measure procedure," page
2. Commands must be sent in order given.
3. Instrument must be addressed to talk after :READ? to acquire data.
Pulse-measure considerations
Measurement speed
Measurement speed for the Pulse Mode can be set from 0.004 to 0.100 PLC. If the speed
setting while in the DC Mode is >0.10 PLC, it will change to 0.004 when the Pulse Mode
is selected. If the speed setting is <0.10 PLC, that setting will be retained when the Pulse
Mode is selected. If the Pulse Mode speed setting is <0.01, it will change to 0.01 when the
DC Mode is selected.
For remote operation, the parameter value for the [:SENSe] ... :NPLC <b> commands is
limited to 0.10. Details are provided in
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
2400 Series SourceMeter
"Front panel pulse-measure procedure," page
2,3
Commands
*RST
Restore GPIB defaults.
:SOUR:FUNC:SHAP PULS
Select Pulse Mode.
:SOUR:PULS:WIDT 0.002
Set pulse width to 2msec.
:SOUR:PULS:DEL 0.003
Set pulse delay to 3msec.
:SENS:VOLT:NPLC 0.08
Set speed to 0.08 PLC.
:TRIG:COUN 25
Set to output 25 pulses.
:SOUR:FUNC VOLT
Select V-Source.
:SOUR:VOLT:MODE FIXED
Fixed voltage source mode.
:SOUR:VOLT:RANG 20
Select 20V source range.
:SOUR:VOLT:LEV 10
Source output = 10V.
:SENS:CURR:PROT 10E-3
10mA compliance.
:SENS:FUNC "CURR"
Current measure function.
:SENS:CURR:RANG 10E-3
10mA measure range.
:READ?
Output 25 pulses, acquire
25 pulse readings.
5-12.
Section
7, "Speed."
®
User's Manual
5-12.