Table 3-2 Compliance Limits - Keithley SourceMeter 2400 User Manual

Sourcemeter 2400 series source measure unit
Hide thumbs Also See for SourceMeter 2400:
Table of Contents

Advertisement

3-6
Basic Source-Measure Operation
Compliance limit
When sourcing voltage, the SourceMeter can be set to limit current. Conversely, when
sourcing current, the SourceMeter can be set to limit voltage. The SourceMeter output will
not exceed the compliance limit.
range. See
Section 6
Table 3-2
Compliance limits
2400
Maximum
Measure
compliance
range
value
200mV
±210mV
2V
±2.1V
20V
±21V
200V
±210V
1µA
±1.05µA
10µA
±10.5µA
100µA
±105µA
1mA
±1.05mA
10mA
±10.5mA
100mA
±105mA
1A
±1.05A
2425/2430
Maximum
Measure
compliance
range
value
200mV
±210mV
2V
±2.1V
20V
±21V
100V
±105V
10µA
±10.5µA
100µA
±105µA
1mA
±1.05mA
10mA
±10.5mA
100mA
±105mA
1A
±1.05A
3A/10A
*
* ±3.15A (2425 and 2430 DC mode)
±10.5A (Pulse mode; 2430 only)
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
2400 Series SourceMeter
Table 3-2
summarizes compliance limits according to
for more details on compliance limit.
2410
Maximum
Measure
compliance
range
value
200mV
±210mV
2V
±2.1V
20V
±21V
1000V
±1.1kV
1µA
±1.05µA
10µA
±10.5µA
100µA
±105µA
1mA
±1.05mA
20mA
±21mA
100mA
±105mA
1A
±1.05A
2440
Maximum
Measure
compliance
range
value
200mV
±210mV
2V
±2.1V
10V
±10.5V
40V
±42V
10µA
±10.5µA
100µA
±105µA
1mA
±1.05mA
10mA
±10.5mA
100mA
±105mA
1A
±1.05A
5A
±5.25A
®
User's Manual
2420
Maximum
Measure
compliance
range
value
200mV
±210mV
2V
±2.1V
20V
±21V
60V
±63V
10µA
±10.5µA
100µA
±105µA
1mA
±1.05mA
10mA
±10.5mA
100mA
±105mA
1A
±1.05A
3A
±3.15A

Advertisement

Table of Contents
loading