Basic Source-Measure Operation
The full range of operation is explained in
ating
boundaries."
NOTE
Output transient recovery — The time required for the V-source to recover to its
original value (within 0.1% plus load regulation errors) after a step change in
load current is <250µsec. This does not include the response time of autorang-
ing or the second order effects on loads that are not purely resistive.
Load regulation — The voltage specification for V-source mode load changes is
0.01% +1mV. This means that on the 200mV range, the load current can be
changed from zero to full scale with less than 1.02mV of error. Calculation:
error = (0.01% x 0.2V) + 1mV = 1.02mV
Assuming a 0 to 1A change in current, the output impedance equates to 1.02m Ω
(1.02mV/1A = 1.02m Ω ). This level can only be achieved using 4-wire remote
sensing.

Table 3-1

Source-measure capabilities
2400
Range
Source
200mV
±210mV
2V
±2.1V
20V
±21V
200V
±210V
1µA
±1.05µA
10µA
±10.5µA
100µA
±105µA
1mA
±1.05mA
10mA
±10.5mA
100mA
±105mA
1A
±1.05A
Max Power = 22W
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
2400 Series SourceMeter
Section
6,
"Overheating
2410
Measure
Range
Source
Measure
200mV
±211mV
±210mV
±211mV
±2.11V
2V
±2.1V
±2.11V
±21.1V
20V
±21V
±21.1V
±211V
1000V
±1.1kV
±1.1kV
±1.055µA
1µA
±1.05µA
±1.055µA
±10.55µA
10µA
±10.5µA
±10.55µA
±105.5µA
100µA
±105µA
±105.5µA
±1.055mA
1mA
±1.05mA
±1.055mA
±10.55mA
20mA
±21mA
±21.1mA
±105.5mA
100mA
±105mA
±105.5mA
±1.055A
1A
±1.05A
±1.055A
Max Power = 22W
®